LC-2030C NT - 規格

體型液相層析儀

Pump

Degassing unit

Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 µL)

Pumping method

Parallel-type double plunger

Pulsation

≤ 0.1 MPa (1.0 mL/min,10 MPa, Water)

Flow rate setting range

0.0001 to 10 mL/min

Flow rate accuracy

≤ ±1% or ±2 µL/min, below whichever is greater (0.01 to 2 mL/min, Specified conditions)
≤ ±2% (2 to 5 mL/min, Specified conditions)

Flow rate precision

≤ 0.06 %RSD or ≤ 0.02 minSD, below whichever is greater

Configuration

Four-solvent low-pressure gradient

Gradient / range of set concentrations

0 to 100%, in 0.1% steps

Gradient / concentration accuracy

±0.5% (0.1 to 2 mL/min, 1 to 20 MPa, Specified conditions)

Gradient / concentration precision

±0.1% (1 mL/min, 10 MPa, Specified conditions)

Maximum pressure

Packed column side : 34 MPa (0.0001 to 5 mL/min) , 22 MPa (5.0001 to 10 mL/min)
Slide-in column side : 7 MPa (0.0001 to 0.3 mL/min)

System Delay Volume

approximately 615 µL (Packed column side)
approximately 620 µL (Slide-in column side) (Option: none)

顯示更多
顯示較少

Autosampler

Injection method

Total-volume sample injection

Injection volume accuracy

±1% (50 µL, N=10)

Injection volume setting range

0.1 to 100 µL (Option:0.1 to 50 µL)

Injection volume reproducibility

RSD ≤0.20% (5.0 to 100 µL), RSD ≤0.25% (2.0 to 4.9 µL)
RSD ≤0.5% (1.0 to 1.9 µL), RSD ≤1.0% (0.5 to 0.9 µL)

Cross-contamination

0.0025% (Caffeine, Specified conditions)

Injection cycle time

Min. 14 sec (Specified conditions)

Samples for processing

336 (1 mL), 216 (1.5 mL), 112 (4 mL), 4 (MTP/DWP)

Sample cooler

4 to 45°C

Injection linearity

≥0.9999 (1 to 100 µL, Specified conditions)

顯示更多
顯示較少

Column Oven

Heating and cooling method

Forced air circulation method

Containable column size

Slide-in column
Packed column (1 pieces at 10 to 25 cm)

Temperature control range

Room temperature −5 to 60°C,
Setting range 4 to 60°C (To 50°C for slide-in column)

Temperature control precision

±0.1°C

Temperature stability

±0.8°C (Specified conditions)

Flow rate switching valve

Not available

顯示更多
顯示較少

PDA Detector

Wavelength range

190 to 800 nm

Spectral resolution

1.4 nm (Specified condition)

Slit width

1.2 nm, 8 nm

Device resolution

0.6 nm/pixel

Number of photodiode array elements

1024

Wavelength accuracy

≤ ±1 nm

Noise level

≤ ±3 × 10−6 AU (250 nm, reference: 350 nm, Specified conditions)

Drift

≤ 500 × 10−6 AU/h (250 nm, reference: 350 nm, Specified conditions)

Linearity

Up to 2 AU (5%)

Sampling rate

Up to 100 Hz

Light source

Deuterium (D2) lamp (Standard), tungsten (W) lamp (option)

Flow Cell

8 µL (10mm, TC), 12 MPa

Option cell

Conventional: 10 µL (10 mm, TC),
Semi-micro: 2.5 µL (5 mm, TC)

顯示更多
顯示較少

Miscellaneous

Dimensions

W410×H605×D500 mm

重量

67 kg

Available pH range

1 to 13

Materials for parts in contact with liquids

Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby,
sapphire, Hastelloy® C, GFP, ceramic, PFA, quartz, PPS

Workstation

LabSolutions LC/GC Ver. 5.92 or later,
LabSolutions DB/CS Ver. 6.82 SP1 or later (Incompatible with LCsolution )

顯示更多
顯示較少
{"title":"\u898f\u683c\u6587\u4ef6\u4e0b\u8f09","description":"\u4e0b\u8f09\u6700\u65b0\u898f\u683c\u6587\u4ef6","source":"product","key":4575,"max":"30","filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"\u4e0b\u8f09","description":"\u4e0b\u8f09\u6700\u65b0\u578b\u9304","source":"product","key":4575,"max":"30","filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}